Description
DYW-3A
Characteristics
1=>To be used in production of medicines, dye-stuffs, perfumes and other organic crystal substances.
2=>Observing the process through visual window or choose auto mode to measure.
3=>Use PID to control the temperature.
4=>It can measure 3 samples at the same time and calculate the average automatically.
5=>Color Touch Screen.
Specifications
|
Parameters/Model DYW-3A |
|
|
Measuring range |
Room temperature -360 °C |
|
Measuring mode |
Auto / Visual |
|
Resolution |
0.1 °C |
|
Linear heating-up rate |
0.1°C/min -10°C/min |
|
Accuracy |
≤ 200°C:±0.5°C ;>200°C:±0.8°C |
|
Repeatability |
0.3°C(heating rate: 1.0°C/min) |
|
Linear heating rate error |
±10% |
|
Number of capillary |
3 |
|
Capillary dimensions |
Outer diameter: φ1.4mm Inner diameter: φ1.0mm Length: 90mm |
|
Height of sample |
3mm |
|
Display |
5.6 inch color touch screen |
|
Calibration method |
Multi-point calibration |
|
Data storage |
100 sets |
|
Interface |
U disk,USB,RS232 |
|
Power supply |
220V±22 V,50Hz±1 Hz,180w |
|
Package Dimension |
500mm×430mm×370mm |
|
Weight(Gross) |
13kg |